THE CESIUM BEAM FREQUENCY STANDARD NRLM-II
Y. Koga, Y. Nakadan et J. YodaNational Research Laboratory of Metrology, Tsukuba Science City, Japan.
Abstract
Design and performance of the cesium beam frequency standard, NRLM-II, are reported. The interaction length 1 m is adopted from the viewpoints of the C-field homogeneity and the signal to noise ratio of resonance signal. The frequency stability is estimated as better than 8X10-12τ-1/2 for short term and 8X10-14 for 1˜2 days. Preliminary evaluation of accuracy is about 5X10-13, and the standard frequency is estimated as +(2˜3±5)X10-13 higher than that of TAI.