TESTS OF MULTILAYERS BASED SOFT X-RAY MONOCHROMATORS FOR SYNCHROTRON RADIATION
M. GRIONI1, 2, F . SCHAEFERS3, J.B. GOEDKOOP1, J.C. FUGGLE1, J.L. WOOD4 and H. van BRUG51 RIM, University of Nijmegen, NL-6525 ED Nijmegen, The Netherlands
2 LURE, Université de Paris-Sud, Bât. 209D, F-91405 Orsay Cedex, France
3 Berliner Elektronenenspeicherring-Gesellschaft für Synchrotronstrahlung GmbH (BESSY), D-1000 Berlin 33, F.R.G.
4 Ovonic Synthetic Materials, Troy, MI 48084, U.S.A.
5 FOM, Institute for Plasma Physics, NL-3430 Nieuwegein, The Netherlands
Abstract
We used multilayer/KAP and multilayer/multilayer combinations in a double-crystal monochromator at the BESSY storage ring. Throughput and resolution of the ML/KAP combination compare favourably with known sources for spectroscopic applications in the 600-800 eV energy range. The ML/ML high throughput appears suited for trace-element analysis and fluorescence or Auger experiments that do not require a highly monocromatic light. These results suggest the use of ML/ML prefilters in high-resolution, multi-element soft X-ray monochromators.