A MULTI-CHANNEL PHOTOELECTRIC SPECTROMETER EMPLOYING A FABRY-PEROT ETALON AND AXICON

G. I. KATCHEN, J. KATZENSTEIN and L. LOVISETTO

Culham Laboratory, Culham, Abingdon, Berks, England


Abstract
The principle of operation, design and construction of a multi-channel photoelectric spectrometer using a Fabry-Perot etalon and an axicon is described in detail. The axicon maps the resolution intervals of the etalon from concentric rings of constant area in the focal plane of the objective into circular areas located at various points along the optic axis. The light is collected from these areas by rectilinear light guides which are optically coupled to flexible fiber bundles leading to individual photomultipliers. The number of channels is limited only by the available etalon finesse. The mean inclination of the order of interference is arbitrary ; the resolution intervals in the vicinity of normal incidence need not be used.


Résumé
Le principe de construction et de fonctionnement d'un spectromètre photoélectrique multicanal utilisant un étalon Fabry-Perot et un axicon est décrit en détail. Les anneaux, de surface constante, recueillis dans le plan focal de l'objectif, sont transformés par l'axicon en surfaces circulaires placées en différents points de l'axe optique. Chaque surface est reliée à un photomultiplicateur par l'intermédiaire d'un guide optique rectiligne couplé à une fibre flexible. Le nombre des canaux n'est limité que par la finesse de l'étalon et le rayon moyen des anneaux est arbitraire. Enfin les anneaux proches du centre n'ont pas besoin d'être utilisés.